EasyManuals Logo

Keithley 2700 User Manual

Keithley 2700
500 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
Model 2700 Multimeter/Switch System User’s Manual Basic DMM Operation 3-53
d. TEMPerature:TCouple:RJUNction:SIMulated <n> [, <clist>]
The units for the simulated reference temperature depend on the present
temperature measurement units as set by UNIT:TEMPerature (see Ref h).
NOTE The following command can instead be used to set the simulated
reference temperature:
TEMPerature:RJUNction:SIMulated <n> [, <clist>]
e. TEMPerature:FRTD:RZERo <NRf> [, <clist>]
TEMPerature:FRTD:ALPHa <NRf> [, <clist>]
TEMPerature:FRTD:BETA <NRf> [, <clist>]
TEMPerature:FRTD:DELTa <NRf> [, <clist>]
These commands are used to set the parameters for the USER RTD type. Note
that the RZERo command sets the “
Ω at 0°C” parameter. When any of these
commands are sent, the USER RTD type is automatically selected.
f. FREQuency:THReshold:VOLTage:RANGe <n> [, <clist>]
PERiod:THReshold:VOLTage:RANGe <n> [, <clist>]
These commands are used to specify the expected input level. The instrument
will then automatically select the most sensitive current or voltage threshold
range.
g. FREQuency:APERture <n> [, <clist>]
PERiod:APERture <n> [, <clist>]
The rate annunciators indicate the following aperture settings:
SLOW = 1 sec
MED = 0.1 sec
FAST = 0.01 sec
For all other aperture times, the rate annunciators are turned off.
h. UNIT:TEMPerature <name>
To set temperature measurement units to °C, use the C or CEL parameter. For
°F, use the F or FAR parameter. For Kelvin, use the K parameter.
i. Trigger and retrieve readings:
NOTE Detailed information on the processes to trigger and retrieve
readings is provided in
Section 8, Section 13, and Appendix D.
INITiate:CONTinuous <b>
INITiate
With continuous initiation disabled (INITiate:CONTinuous OFF), you can use
the INITiate command to trigger one or more measurements.
NOTE Note that sending INITiate while the instrument is performing
measurements will cause error -213 (init ignored).

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD

Related product manuals