Model 2700 Multimeter/Switch System User’s Manual Triggering 8-15
•*RST
What happens next depends on the state of initiation. If continuous initiation is already
enabled, the instrument will leave the idle state. SYSTem:PRESet enables continuous
initiation. Therefore, operation will immediately leave the idle state when it is sent. The
*RCL command will do the same if INITiation:CONTinuous ON is a user-saved default.
*RST disables continuous initiation. Therefore, the instrument will remain in the idle
state.
Either of the following two initiate commands will take the instrument out of the idle state:
•INITiate
INITiate:CONTinuous ON
NOTE While in remote, pressing the LOCAL key restores continuous front panel
operation.