Model 2700 Multimeter/Switch System User’s Manual Triggering 8-15
•*RST
What happens next depends on the state of initiation. If continuous initiation is already
enabled, the instrument will leave the idle state. SYSTem:PRESet enables continuous
initiation. Therefore, operation will immediately leave the idle state when it is sent. The
*RCL command will do the same if INITiation:CONTinuous ON is a user-saved default.
*RST disables continuous initiation. Therefore, the instrument will remain in the idle
state.
Either of the following two initiate commands will take the instrument out of the idle state:
•INITiate
INITiate:CONTinuous ON
NOTE While in remote, pressing the LOCAL key restores continuous front panel
operation.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD