8-16 Triggering Model 2700 Multimeter/Switch System User’s Manual
Figure 8-10
Trigger model (remote operation)
Control
Source
Event
Detection
No
START
:INIT (:IMM)
or
:INIT:CONT ON
?
Yes
:ABOrt
*RCL 0
:SYST:PRES
*RST
:INIT (:IMM)
or
:INIT:CONT ON
?
No
Yes
Idle
and
Initiate
:Trigger:Source Immediate
:Trigger:Source External
:Trigger:Source Timer
:Trigger:Source Manual
:Trigger:Source BUS
:Trigger:Delay <n>
:Trigger:Delay:AUTO <b>
Another
Trigger
?
Output
Trigger
Another
Sample
?
Device
Action
Yes
No
Yes
No
:Trigger:Signal
:Sample:Count <n>
:Trigger:Count<n> | INFinity
Timer >
Delay
?
Timer
Enabled
?
Yes
No
Timer
Timer
Bypass
Yes
No
Delay
(Auto or Manual)

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD