15-22 SCPI Reference Tables Model 2700 Multimeter/Switch System User’s Manual
SYSTem
:CARDX
[:STARt]? Request lowest numbered volts/2-wire channel
(usually 1); 0 = voltage measurements not
supported.
:END? Request highest numbered volts/2-wire channel.
0 = voltage measurements not supported.
:ACHannel Path to query amps channels:
[:STARt]? Request lowest numbered amps channel;
0 = amps measurements not supported.
:END? Request highest numbered amps channel;
0 = amps measurements not supported.
:ICHannel Path to query isolated channels. An isolated
channel includes 2/4-pole and backplane
relays.
[:STARt]? Request the first isolated channel.
0 = no isolated channels.
:END? Request the last isolated channel.
0 = no isolated channels.
:AOUTput Path to query analog output channels:
[:STARt]? Request highest numbered analog output
channel; 0 = analog output not supported.
:END? Request lowest numbered analog output
channel; 0 = analog output not supported.
:DOUTput Path to query digital output channels:
[:STARt]? Request lowest numbered digital output
channel; 0 = digital output not supported.
:END? Request highest numbered digital output
channel; 0 = digital output not supported.
:TCHannel? Query the totalizer channel.
:DINPut Path to query digital input channels:
[:STARt]? Request lowest numbered digital input port;
0=digital input not supported.
:END? Request highest numbered digital output
channel; 0 = digital input not supported.
:SNOpen? Query whether card is “single no-open” type
(i.e., 7711): 1 = yes, 0 = no.
Table 15-7 (continued)