Model 2700 Multimeter/Switch System User’s Manual Close/Open Switching Module Channels 2-13
c. ROUTe:CLOSe?
This query command returns a <clist> of closed measurement channels, including
paired channels for 4-wire functions.
This query command will not return non-measurement channels, such as
backplane isolation channels and the pole-mode channel.
d. ROUTe:OPEN:ALL
This command functions the same as the front panel OPEN key (ALL menu
option). It simply opens all channels (including non-measurement channels)
installed in the mainframe.
Remote programming example (system channel operation)
The following example assumes a Model 7700 installed in slot 1, and the Ω4 function of
the Model 2700 is selected. This command sequence connects channel 101 and its paired
channel (111) to DMM Input and Sense as shown in
Figure 2-2.
ROUT:OPEN:ALL ' Open all channels.
ROUT:CLOS (@101) ' Close channels 101, 111, 123, 124, and 125.