100 Keysight N9010A EXA Service Guide
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; Normalize
is not available while Trace
Math is on
The Normalize function works by doing trace manipulation. So if
trace math is on you cannot turn on normalization.
-221 Settings conflict; Option
not available
You have attempted to perform an action for which a required
option is not installed
-221 Settings conflict; Param
only available when DUT is a
freq converter
SCPI only message. The sideband and freq context parameters are
only available when a freq conversion setup is in use. Change
setup to contain a freq conversion to use these parameters.
-221 Settings conflict; Param
only available when External
LO Mode is Swept
SCPI only message. This parameter is only available when the LO
mode is set to Swept. Change the LO Mode to Swept.
-221 Settings conflict; Param
only available when External
LO Mode is Fixed
SCPI only message. This parameter is only available when the LO
mode is set to Fixed. Change the Freq Mode to Fixed.
-221 Settings conflict; Param
only available when
Frequency Mode is Fixed
SCPI only message. This parameter is only available when the Freq
mode is set to Fixed. Change the Freq Mode to Fixed.
-221 Settings conflict; Param
only available when
Frequency Mode is Swept
SCPI only message. This parameter is only available when the Freq
mode is set to Swept. Change the Freq Mode to Swept
-221 Settings conflict; Param
only available when valid
cal data exists
SCPI only message. The ‘Apply Calibration’ parameter is only
available when the stored cal data matches the current setup.
Perform a fresh ‘Calibrate Now’ or change setup such that current
cal data is valid.
-221 Settings conflict; Parameter
currently is disabled
This parameter is grayed out (unavailable) in the current context.
Check the individual parameter help/documentation for more
information.
-221 Settings conflict; Preamp
gain is not available in
this Mode
Preamp gain correction is not available in some Modes or
Measurements
-221 Settings conflict; Preamp
unavailable with electronic
attenuator on
The electronic attenuator is on. Internal preamp cannot be used
while you are using the electronic attenuator.
-221 Settings conflict; Reference
marker must be in same
window
A delta marker and its reference must be in the same window. This
error occurs when you try to turn on a delta marker who’s
reference is in a different window.
-221 Settings conflict; Res BW
cannot be auto-coupled while
in Zero Span
The resolution bandwidth cannot be set to auto while you are in
zero span (time domain).
-221 Settings conflict; Scale
Type = Lin is not available
when Normalize is on
Only the Log amplitude scale is available in Normalize, since the
results are always presented as a dB ratio.