Keysight N9010A EXA Service Guide 107
Instrument Messages
-200 to -299, Execution Errors
-224 Illegal parameter value;
Illegal identifier
<identifier>. This value
may already be in use.
The value for the LXI LAN identifier parameter must be unique (i.e.
LAN0 and LAN7 must have different identifier strings).
-224 Illegal parameter value;
Index out of range
When querying the LXI Event Log or the Servo Log, an index may be
used to look at a specific entry. This error occurs if the index provided
does not point to a valid entry.
-224 Illegal parameter value;
Invalid list length
You are trying to set some list measurement settings, but the multiple
lists that you sent were not all the same length. The number of
settings must be consistent from list to list.
-224 Illegal parameter value;
LXI Event <event> already
exists.
This error occurs when you try to add an LXI Event that has already
been added.
-224 Illegal parameter value;
LXI Event <event> contains
illegal characters.
When a new LXI Event is created, it may not use the comma,
semicolon, or newline characters. All other printable ASCII characters
are valid.
-224 Illegal parameter value;
LXI Event <event> does not
exist.
The requested event has not been added yet.
-224 Illegal parameter value;
Measurement not available
You tried to turn on a measurement that is not available in the current
mode.
-224 Illegal parameter value;
trace points not available
for tier
You tried to import a trace file that contains trace points that don't
match the number of trace points available for your software license
tier.
-225 Out of memory There is not enough memory to perform the requested operation.
-225 Out of memory; Memory limit
caused Data Acquisition to
be truncated
-226 List not same length You are using the LIST structure, but have individual lists that are not
the same lengths.
-230 Data corrupt or stale; A legal data element was found, but it could not be used because the
data format or the data structure was not correct. Maybe a new
measurement had been started but had not completed.
-230 Data corrupt or stale;
Measurement data is not
available
Measurement data not available. The measurement that you are trying
to get data from must be the current active measurement. Maybe you
have not initiated the measurement, or it has not completed all the
sweeps/averages needed.
-230 Data corrupt or stale;
Trace contains no data.
Trace cannot be displayed because currently there is no data assigned
to it. Use the functions under the Trace menu, or load a previously
saved trace, to assign data to the trace.
-230 Data corrupt or stale;
Unable to load state from
file
There is something wrong with the state data in the desired file. Maybe
the file is corrupt, or it is from an instrument/version that is not
recognized by the current instrument.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals