Keysight N9010A EXA Service Guide 111
Instrument Messages
-200 to -299, Execution Errors
-278 Macro header not found Indicates that a syntactically legal macro label in the *GMC? query
could not be executed because the header was not previously defined.
-280 Program error There was an execution error in a down-loaded program. The exact
problem cannot be specifically identified.
-281 Cannot create program Indicates that an attempt to create a program was unsuccessful. A
reason for the failure might include not enough memory.
-282 Illegal program name The name used to reference a program was invalid; for example,
redefining an existing program, deleting a nonexistent program, or in
general, referencing a nonexistent program.
-283 Illegal variable name An attempt was made to reference a nonexistent variable in a program.
-284 Program currently running Certain operations dealing with programs may be illegal while the
program is running; for example, deleting a running program might not
be possible.
-285 Program syntax error Indicates that a syntax error appears in a downloaded program. The
syntax used when parsing the downloaded program is device-specific.
-286 Program runtime error
-290 Memory use errors
-291 Out of memory
-292 Referenced name does not
exist
-293 Referenced name already
exists
-294 Incompatible type Indicates that the type or structure of a memory item is inadequate