110 Keysight N9010A EXA Service Guide
Instrument Messages
-200 to -299, Execution Errors
-254 Media full A legal command/query could not be executed because the media
was full
-255 Directory full A legal command or query could not be executed because media
directory was full.
-256 File name not found; A legal command or query could not be executed because the file
name was not found in the specified location.
-257 File name error; A legal command or query could not be executed because there was
an error with the file name on the device media. For example, maybe
you tried to copy to a duplicate file name.
-257 File name error; Allowable
extension is .csv
You are using the wrong type of file extension for the current data/file
type.
-257 File name error; Allowable
extension is .png
You are using the wrong type of file extension for the current data/file
type.
-257 File name error; Allowable
extension is .state
You are using the wrong type of file extension for the current data/file
type.
-257 File name error; Invalid
file name
The filename, directory name, or volume label syntax is incorrect.
-257 File name error; name too
long
-258 Media protected A legal command or query could not be executed because the media
was protected. For example, the write-protect was set
-260 Expression error An error was found with an expression type of data element. The exact
problem cannot be specifically identified.
-261 Math error in expression An expression that has legal syntax could not be executed because of
a math error. For example, maybe you are dividing by zero.
-270 Macro error Indicates that a macro-related execution error occurred.
-271 Macro syntax error Indicates a syntax error within the macro definition
-272 Macro execution error Indicates that a syntactically legal macro program data sequence
could not be executed due to some error in the macro definition
-273 Illegal macro label Indicates that the macro label defined in the *DMC command was a
legal string syntax, but could not be accepted
-274 Macro parameter error Indicates that the macro definition improperly used a macro
parameter placeholder
-275 Macro definition too long Indicates that a syntactically legal macro program data sequence
could not be executed because the string or block contents were too
long for the device to handle
-276 Macro recursion error Indicates that a syntactically legal macro program data sequence
could not be executed because the device found it to be recursive
-277 Macro redefinition\ not
allowed
Indicates that a syntactically legal macro label in the *DMC command
could not be executed because the macro label was already defined

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals