Keysight N9010A EXA Service Guide 109
Instrument Messages
-200 to -299, Execution Errors
-250 Mass storage error; Failed
to Load trace. Bad file
format.
The load trace operation could not be completed, as the input file was
not in the expected format. You can only load traces that were
previously saved using the 'Save Trace' feature.
-250 Mass storage error; File
<filename> wrong type
Attempt to import a data file that is not the proper type for this
operation.
-250 Mass storage error; File
<filename> and instrument
version mismatch
While opening a file, there was a mismatch between file version or
model number with instrument version or model number. The import
still tried to load as much as possible, but you should check it closely.
-250 Mass storage error; File
contains incorrect data for
this operation
There is a mismatch between the file data type of the file specified and
the destination indicated. For example, a correction set cannot be
loaded/imported into a limit line.
-250 Mass storage error; File
empty
Cannot save trace because it contains no data. Check that the trace is
turned on and contains some valid data.
-250 Mass storage error; Invalid
register number for *SAV or
*RCL
You have used the *SAV command to save a state to a non existent
state register.
Or
You have used the *RCL command to recall a state register that wasnt
previously saved with the *SAV command.
-250 Mass storage error; Lock
violation
The process cannot access the file because another process has
locked a portion of the file.
-250 Mass storage error; No file
names available
Attempt to use the auto file name generation when all 10,000 file
names are taken.
-250 Mass storage error; Open
failed
The system cannot open the device or file specified. This could be
because the storage media is full, or possibly due to a filename error.
If using an external storage devise, check that the devise is properly
formatted.
-250 Mass storage error; Read
fault
The system cannot read from the specified device.
-250 Mass storage error;
Register <number> empty
Attempt to recall a register with nothing in it
-250 Mass storage error; Sharing
violation
The process cannot access the file because it is being used by another
process.
-250 Mass storage error; Too
many open files
The system cannot open the file.
-250 Mass storage error; Write
fault
The system cannot write to the specified device.
-252 Missing media A legal command or query could not be executed because missing
media.
-253 Corrupt media A removable media was found to be bad or incorrectly formatted. Any
existing data on the media may have been lost.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals