Keysight N9010A EXA Service Guide 609
Post-Repair Procedures
Post-Repair Procedures
Enabling Spread Spectrum
1. Press the Left Arrow on the external keyboard to highlight the Advanced
tab as shown in Figure 16-9.
Figure 16-9 Advanced Tab Highlighted
2. Press the Down Arrow on the external keyboard to highlight CK-408
Spread Spectrum
3. Use the “+” key on the external keyboard to change the setting from
[Disabled] to [Enabled]
4. Press “F10” on the keyboard to save changes and exit.