Keysight N9010A EXA Service Guide 641
Functional Tests
Frequency Readout Accuracy
7. Repeat step 4 through step 6 until the Marker Frequency Readout column
of Table 17-2 is complete.
Table 17-2 Frequency Readout Accuracy Results
Span
(MHz)
Center
Frequency (MHz)
Minimum
Marker Frequency
Readout
Maximum
2990 1505 1495.9 MHz 1514.1 MHz
127.2 1505 1504.56 MHz 1505.44 MHz
54.1 1505 1504.8122 MHz 1505.1878 MHz
7.95 1505 1504.97240 MHz 1505.0276 MHz
0.106 1505 1504.999630 MHz 1505.000370 MHz
1.98 517.59 517.58316 MHz 517.59684 MHz
1.98 832.50 832.49316 MHz 832.50684 MHz