3 Programming the Analyzer
STATus Subsystem
Mode All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?
Example STAT:QUES:INT:UNC:ENAB 1 Oversweep (Meas Uncal) will be reported to the Integrity Summary of
the Status Questionable register.
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Integrity Uncalibrated Event Query
This query returns the decimal value of the sum of the bits in the Questionable Integrity Uncalibrated Event
register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?
Example STAT:QUES:INT:UNC?
Preset 0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision Prior to A.02.00
Questionable Integrity Uncalibrated Negative Transition
This command determines which bits in the Questionable Integrity Uncalibrated Condition register will set
the corresponding bit in the Questionable Integrity Uncalibrated Event register when the condition register
bit has a negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode All
Remote Command
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition <integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition?
Example STAT:QUES:INT:UNC:NTR 1 Oversweep cleared will be reported to the Integrity Summary of the
Status Questionable register.
EMI Receiver Mode Reference 133