10 Monitor Spectrum Measurement
Meas Setup
Key Path
Meas Setup, Signal List, Mark Signals
Notes Front panel only.
Dependencies Yes
Couplings When the signal list is cleared, this value will reset to 0.
Preset If the signal list is empty, =0
If the signal list is not empty, = number of peaks in signal list table
State Saved Saved in instrument state
Min 0
Max Number of peaks in signal list table.
Initial S/W Revision A.13.00
Mark Signal
Marks the selected signal.
Key Path
Meas Setup, Signal List, Mark Signals
Remote Command
:CALCulate:SLISt:MARK:SIGNal <integer>
Example CALC:SLIS:MARK:SIGN 4 Mark signal #4.
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
If the selected signal is not in the list, an error is generated and this function cannot be performed.
Initial S/W Revision A.13.00
Mark Duplicates
This menu enables you to mark the signals that are duplicates. A signal is considered to be a duplicate if it
is within the width of the 3 dB RBW value. Signals cannot be duplicate unless they are measured using the
same 3 dB RBW.
Key Path
Meas Setup, Signal List, Mark Signals
Initial S/W Revision A.13.00
All
Marks all duplicate signals.
Key Path
Meas Setup, Signal List, Mark Signals, Mark Duplicates
Remote Command
:CALCulate:SLISt:MARK:DUPLicates[:ALL]
Example CALC:SLIS:MARK:DUPL
Notes If the signal list contains no data, an error is generated and this function cannot be performed.
Initial S/W Revision A.13.00
782 EMI Receiver Mode Reference