6 System Functions
System
Example :CAL:RFPS:ALER OFF
Notes For model N9038A only.
Error Condition will be generated when the alert is On and any of the RF Preselector alignments has
expired.
Preset This is unaffected by Preset, but is set to ON on a “Restore System Defaults->Align”.
State Saved No
Initial S/W Revision A.08.00
Schedule Setup
Enables you to schedule a task to run automatically at the background based on the recurrence and time
set in the scheduler. Make sure that the Instrument’s local time is accurate as the Scheduler relies on this
information to execute the task.
Key Path
System, Alignments, RF Preselector
Initial S/W Revision A.08.00
Task
There are 3 task that can be selected for the scheduler to run.
Task 1 is the 20 Hz to 30 MHz alignment
Task 2 is the 30 MHz to 3.6 GHz alignment
Task 3 is the 20 Hz to 3.6 GHz alignment.
Key Path
System, Alignments, RF Preselector, Schedule Setup
Mode All
Remote Command
:CALibration:RFPSelector:SCHeduler:TASK T1|T2|T3
:CALibration:RFPSelector:SCHeduler:TASK?
Example :CAL:RFPS:SCH:TASK T1
Notes Changing the task will not reset the Scheduler time and the alignment is based on the current
scheduled configuration to occur.
For model N9038A only.
Preset This is unaffected by Preset but is set to T3 on a “Restore System Defaults->Align”.
State Saved No
Range Task 1 | Task 2 | Task 3
Initial S/W Revision A.08.00
328 EMI Receiver Mode Reference