9 Strip Chart Measurement
Marker
Notes A marker may be placed on a blanked and/or inactive trace, even though the trace is not visible
and/or updating.
An application may register a trace name to be displayed on the key instead of a trace number.
Couplings The state of Marker Trace is not affected by the Auto Couple key.
If a Marker Trace is chosen manually, Auto Init goes to Off for that marker.
Sending the remote command causes the addressed marker to become selected.
Preset 1
State Saved The Marker Trace and state of Auto Init for each marker is saved in instrument state.
Min 1
Max 3
Readback line Trace N where N is the trace number to which the marker is currently assigned.
Initial S/W Revision A.07.00
Couple Marker
When this function is invoked, moving any marker causes an equal X Axis movement” of every other
marker which is active. By equal X Axis movement” we mean that the difference between each marker’s X
Axis value (in the fundamental x-axis units of the trace that marker is on) is preserved, as is the X Axis value
of the marker being moved (in the same fundamental X-axis units).
This may result in markers going off screen.
Key Path
Marker
Remote Command
:CALCulate:SCHart:MARKer:COUPle[:STATe] ON|OFF|1|0
:CALCulate:SCHart:MARKer:COUPle[:STATe]?
Example CALC:SCH:MARK:COUP ON
CALC:SCH:MARK:COUP?
Preset OFF
State Saved Saved in instrument state
Initial S/W Revision A.07.00
All Marker Off
Turns all markers Off.
Key Path
Marker
Remote Command
:CALCulate:SCHart:MARKer:AOFF
638 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals