10 Monitor Spectrum Measurement
AMPTD Y Scale
Dependencies Grayed out if an Amplitude Correction with an Antenna Unit is ON.
Readback dBmV
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00
dBmA
Sets the amplitude unit for the selected amplitude scale (log/lin) to dBmA.
Key Path
AMPTD Y Scale, Y Axis Unit
Example UNIT:POW DBMA
Dependencies Grayed out if an Amplitude Correction with an Antenna Unit is ON.
Readback dBmA
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00
W
Sets the amplitude unit for the selected amplitude scale (log/lin) to watt.
Key Path
AMPTD Y Scale, Y Axis Unit
Example UNIT:POW W
Dependencies Grayed out if an Amplitude Correction with an Antenna Unit is ON.
Readback W
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00
V
Sets the amplitude unit for the selected amplitude scale (log/lin) to volt.
Key Path
AMPTD Y Scale, Y Axis Unit
Example UNIT:POW V
Dependencies Grayed out if an Amplitude Correction with an Antenna Unit is ON.
Readback V
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00
736 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals