10 Monitor Spectrum Measurement
SPAN X Scale
200 kHz 100 Hz 1 kHz
100 kHz 100 Hz 1 kHz
50 kHz 30 Hz 300 Hz
20 kHz 10 Hz 100 Hz
10 kHz 10 Hz 100 Hz
5 kHz 10 Hz 30 Hz
2 kHz 10 Hz 30 Hz
1 kHz 10 Hz 30 Hz
Preset 1 MHz
State Saved Saved in instrument state
Min 1 kHz
Max 10 MHz
Default Unit Hz
Status Bits/OPC
dependencies
non-overlapped
Initial S/W Revision A.13.00
Last Span
Changes the displayed frequency span to the previous span setting. If it is pressed immediately after
Signal Track is turned off, then the span setting returns to the span that was in effect before Signal Track
was turned on.
Key Path
SPAN X Scale
Remote Command
[:SENSe]:MONitor:FREQuency:SPAN:PREVious
Example :MON:FREQ:SPAN:PREV sets the span to the previous value
Initial S/W Revision A.13.00