10 Monitor Spectrum Measurement
Trace/ Detector
Remote Command Notes When TRACE1, the latest contents of trace 1 will be returned.
When TRACE2, the latest contents of trace 2 will be returned.
When TRACE3, the latest contents of trace 3 will be returned.
When SINGle, the current power values of all activated meters are returned separated by commas in
the order meters number.
Backwards
Compatibility SCPI
:TRACe[:DATA]?
Backwards Compatibility
SCPI Notes
This command is included for ESU compatibility. The SCPI is Meas Local and Context Sensitive.
Initial S/W Revision A.13.00
Trace Display (Remote Command Only)
This SCPI sets the state of the trace’s Update state.
Remote Command
DISPlay[:WINDow]:TRACe[1]|2|3[:STATe] ON|OFF|1|0
DISPlay[:WINDow]:TRACe[1]|2|3[:STATe]?
Example DISP:TRAC OFF
DISP:TRAC?
Notes This command is included for ESU compatibility. The suffix [1]|2 and [1]|2|3 are irrelevant to the
measurement, they are added solely for remote language compatibility. This SCPI alias to :TRACe[1]
|2|3:MONitor:UPDate ON|OFF|1|0.
Backwards
Compatibility SCPI
DISPlay:WINDow[1]|2:TRACe[1]|2|3[:STATe]
Initial S/W Revision A.13.00
EMI Receiver Mode Reference 873

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals