8 Frequency Scan Measurement
SPAN X Scale
SPAN X Scale
Displays the menu keys that enable you to control the span parameters. The only function in this menu is
Span, which controls the Span of the display.
Key Path
Front-panel key
Notes Front panel only.
Initial S/W Revision A.07.00
Span
Sets the span of the graticule display.
Key Path
SPAN X Scale
Remote Command
[:SENSe]:FREQuency:SPAN <frequency>
[:SENSe]:FREQuency:SPAN?
Example FREQ:SPAN 140 MHZ
FREQ:SPAN?
Notes Max values are depending on Hardware Options (503, 507, 508, 513, 526)
Dependencies
TheSpan can be limited by Start or Stop Freq limits, if the Midscreen Frequency is such that Start or
Stop hit their limit.
Preset 970 MHz
State Saved Saved in instrument state.
Min 10 Hz
Max Depends on the instrument maximum frequency 10 Hz minimum span. If the knob or step keys are
being used, it depends on the value of other three interdependent parameters Mid-screen Frequency,
Start Frequency and Stop Frequency.
Default Unit Hz
Status Bits/OPC
dependencies
non-overlapped
Initial S/W Revision A.08.00
572 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals