4 Input/Output Functions
Input/Output
Example SENS:CORR:CSET1 ON
Dependencies Changing this from the OFF state to the ON state automatically turns on "Apply Corrections".
Only the first correction array (Correction 1) supports antenna units. When this array is turned on,
and it contains an Antenna Unit other than None”, the Y Axis Unit of the analyzer is forced to that
Antenna Unit. All other Y Axis Unit choices are grayed out.
Note that this means that a correction file with an Antenna Unit can only be loaded into the
Corrections 1 register. Consequently only for Correction 1 does the dropdown in the Recall dialog
include.ant, and if an attempt is made to load a correction file into any other Correction register
which DOES contain an antenna unit, a Mass Storage error is generated.
This command will generate an Option not available” error unless you have the proper option
installed in your instrument.
Preset Not affected by a Preset. Set to OFF by Restore Input/Output Defaults
State Saved Saved in instrument state.
Backwards Compatibility
Notes
Unlike legacy analyzers, Preset does not turn Corrections off (Restore Input/Output Defaults does).
Initial S/W Revision A.02.00
Modified at S/W Revision A.14.00
Properties
Accesses a menu that lets you set the properties of the selected correction.
Key Path
Input/Output, Corrections
Initial S/W Revision A.02.00
Select Correction
Specifies the selected correction. The term "selected correction" is used throughout this document to
specify which correction will be affected by the functions.
Key Path
Input/Output, Corrections
Mode SA, I/Q Analyzer, Phase Noise, VXA, RTSA, EMI Receiver, DVB-T/H, DTMB, DVB-T/H, DTMB, W-
CDMA, LTE & LTE-Adv FDD, LTE & LTE-Adv TDD, Sequence Analyzer, BTooth
Notes The selected correction is remembered even when not in the correction menu.
Preset Set to Correction 1 by Restore Input/Output Defaults
Readback Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction
6|Correction7|Correction8
Initial S/W Revision A.02.00
Modified at S/W Revision x.14.50
EMI Receiver Mode Reference 185

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals