11 APD Measurement
Marker
Example CALC:MARK3:X 30 Turn on maker 3 in Normal mode, by default will set the marker amplitude to 30
dBuV.
CALC:MARK3:X? Turn on maker 3 in Normal mode and query the X axis value.
Notes This SCPI is alias to :CALC:APD:MARK[1]|2|3|4:MODE POS; :CALC:APD:MARK[1]|2|3|4:X?
Preset 25 dBuV
Initial S/W Revision A.13.00
Normal Marker Y Axis Value
Enables you to set the selected marker to Normal mode, and gets the marker YAxis Values.
Remote Command
:CALCulate[1]|2:MARKer[1]|2|...|4:Y?
Example CALC:MARK3:Y? Turn on maker 3 in Normal mode and query the Y axis value.
Notes This SCPI is alias to :CALC:APD:MARK[1]|2|3|4:MODE POS; :CALC:APD:MARK[1]|2|3|4:Y?
Initial S/W Revision A.13.00
Delta Marker State
Enables you to toggle the Delta marker state on/off.
Remote Command
:CALCulate[1]|2:DELTamarker[1]|2|...|4[:STATe] ON|OFF|1|0
:CALCulate[1]|2:DELTamarker[1]|2|...|4[:STATe]?
Example CALC:DELT3 ON Turn on maker 3 in Delta mode.
Notes This SCPI is alias to :CALC:APD:MARK[1]|2|3|4:MODE DELT.
Preset OFF
Initial S/W Revision A.13.00
Delta Marker X Axis Value
Enables you to set the selected marker to Delta mode, and sets/ gets the marker X Axis Values. The
command input, in relative values, refers to the reference marker while the query always returns absolute
values in order to make it compatible with ESU.
Remote Command
:CALCulate[1]|2:DELTamarker[1]|2|...|4:X <value>
:CALCulate[1]|2:DELTamarker[1]|2|...|4:X?
Example CALC:DELT3:X 30 Turn on maker 3 in Delta mode, set the relative delta marker amplitude to 30 dB.
CALC:DELT3:X? Turn on maker 3 in Delta mode and query the X axis value.
Notes This SCPI is alias to :CALC:APD:MARK[1]|2|3|4:MODE DELT; :CALC:APD:MARK[1]|2|3|4:X? returns the
relative values when marker in Delta mode.
Preset 25 dBuV
Initial S/W Revision A.13.00
924 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals