9 Strip Chart Measurement
SPAN X Scale
Min 20 ms
Max 7200 s
Initial S/W Revision A.07.00
Strip Chart Max Duration
Sets the maximum duration for strip chart to record the data.
Key Path
SPAN X Scale
Remote Command
:DISPlay:SCHart:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:MAX:DURation <time>
:DISPlay:SCHart:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:MAX:DURation?
Example DISP:SCH:VIEW:WIND:TRAC:X:MAX:DUR 100 s
DISP:SCH:VIEW:WIND:TRAC:X:MAX:DUR?
Couplings No
Preset 500 s
State Saved Saved in instrument state.
Min 200 ms
Max 1200 s
Initial S/W Revision A.07.00
Full Scale
Sets the Strip Chart display in full scale, this key will set Ref Value to 0s and max Scale/Div.
Key Path
SPAN X Scale
Remote Command
:DISPlay:SCHart:VIEW[1]:WINDow[1]:TRACe:X[:SCALe]:FULL
Example DISP:SCH:VIEW:WIND:TRAC:X:MAX:FULL
Initial S/W Revision A.07.00