10 Monitor Spectrum Measurement
View/Display
Meters RBW Lines
When Meters RBW Lines is turned on, two blue vertical lines appear on the display, equal distances from
the center of the graticule. Their spacing matches the current setting for the Meters RBW as shown below.
If the Meters RBW is wider than the span of the Monitor Spectrum display, these lines are not shown.
Key Path
View Display
Remote Command
:DISPlay:MONitor:RBWLines[:STATe] ON|OFF|1|0
:DISPlay:MONitor:RBWLines[:STATe]?
Example DISP:MON:RBWL ON
DISP:MON:RBWL?
Preset OFF
State Saved Saved in instrument state
Initial S/W Revision A.14.50
Example Views
Below are the views with Expand Meters Off (top) and Expand Meters on (bottom)