9 Strip Chart Measurement
Meas Setup
Meas Setup
Displays the menu keys that enable you to setup the measurement.
Key Path
Front-panel key
Notes Front panel only.
State Saved Saved in instrument state.
Initial S/W Revision A.07.00
Analog Demod Tune & Listen
The Analog Demod Tune & Listen key opens the Analog Demod menu that contains keys to turn the demod
function on and off and select the modulation type.
Key Path
Meas Setup
Remote Command
[:SENSe]:SCHart:DEMod AM|FM|PM|OFF
[:SENSe]:SCHart:DEMod?
Example SCH:DEM AM
SCH:DEM?
Dependencies When Tune & Listen is turned on, only Meter 1 will return the valid measurement result. Meter 2 & 3
will return 1000dBm. Meter 2 & 3 will return valid measurement results when Turn & Listen is
turned off.
Preset OFF
State Saved Saved in instrument state.
Range AM|FM|PM|OFF
Initial S/W Revision A.07.00
AM
Pressing this key, when it is not selected, selects and activates the AM demodulation function.
Key Path
Meas Setup, Analog Demod Tune&Listen
Example SCH:DEM AM Turns AM demodulation function ON.
State Saved Saved in instrument state.
Initial S/W Revision A.07.00
EMI Receiver Mode Reference 655

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals