11 APD Measurement
Meas Setup
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example APD:DEM:FM:DEEM US25
Readback
25 s
Initial S/W Revision A.13.00
50 µs
Sets the De-emphasis time constant to 50 µs.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example APD:DEM:FM:DEEM US50
Readback
50 s
Initial S/W Revision A.13.00
75 µs
Sets the De-emphasis time constant to 75 µs.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example APD:DEM:FM:DEEM US75
Readback
75 s
Initial S/W Revision A.13.00
750 µs
Sets the De-emphasis time constant to 750 µsec.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example APD:DEM:FM:DEEM US750
Readback
750 s
Initial S/W Revision A.13.00
PM
Pressing this key, when it is not selected, selects and activates the PM demodulation function.
Key Path
Meas Setup, Analog Demod Tune&Listen
Example APD:DEM PM Turns PM demodulation function ON.
State Saved Saved in instrument state.
Initial S/W Revision A.13.00
EMI Receiver Mode Reference 955

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals