J Communications Based on Modbus Protocol
196
5. Enter any of the following values to address 56850 to fine-adjust the span.
Setting Fine adjustment
1
Slightly increases the adjustment amount.
10
Increases the adjustment amount.
-1 Slightly decreases the adjustment amount.
-10
Decreases the adjustment amount.
6. Set [1] to address 56848 (Calibration complete).