3 Calibration
49
CAUTION
The measuring weight does not accept [0]. When you select [0] and press ,
flashing appears and then reappears and flashes.
For setting procedures with and , see Section 1.6, "Basic key operation".
4. Press .
flashes.
5. Press .
The currently set weight mass appears.
6. Similar to the case of measuring weight, change the weight mass on an as-needed basis.
CAUTION
The weight mass does not accept the value exceeding the measuring weight or [0].
When you select either value and press , flashing appears and
then reappears and flashes.
7. Press .
appears.
8. Place the load equal to the set weight mass to the measuring unit.
9. Press .
flashes. Then, continues flashing to allow you to return to Measurement
mode.
CAUTION
When the input from the measuring unit exceeds 3.1 mV/V, flashing
appears. When the input is lower than 0.0 mV/V, flashing appears.
These outliers are outside the scope of span calibration.