5 Settings Related to Detection
80
4. Using and , change the numerical value to and press .
The value showing the current setting appears.
Numerical value Setting details
0 Disable
1 Detection result judgment
2 Waveform comparison
0 (Disable)
For setting procedures with and , see Section 1.6, "Basic key operation".
5. Using , set [2] and press .
CAUTION
When [0] (Entire section) or [1] (Specified section) has been set to P function PF-02,
appears.
After [2] (Waveform comparison) is set to PF-50, changing the setting of PF-02 to
[0] (Entire section) or [1] (Specified section) will automatically set PF-50 to [0]
(Disable).
6. Press .
7. After confirming is displayed, press .
The currently set upper limit appears.
99 999
8. Using and , set the upper limit.
You can set any value between -99 999 and 99 999.
9. Press .
10. After confirming is displayed, press .
The currently set lower limit appears.
-99 999
11. Using and , set the lower limit.
You can set any value between -99 999 and 99 999.
12. Press twice to return to Measurement mode.
Default
Default
Default