Chapter 4 I/O SETTINGS
Chapter 4
Settings for I/O
4-27
ZW
User’s Manual
■ GATE signal setting
Sets the output time for the synchronization signal (GATE) for taking in signals 
externally.
As an example, here is an explanation of the procedure for setting the output time to 
"100 ms".
Operating procedure
Setting [Display] Description
GATE period [GATE] The GATE signal is output for the set time.
Steps
Key operation Display Description
1 - 3 For moving to BINARY, see steps 1 to 3 in p.4-22
4 Press   or   keys to select either of "GATE" and 
press  key.
5 Set "100" as the output time.
The current setting value is displayed on the sub-
display.
Press   key to enter the editing mode, and the 
sub-display blinks.
To edit numerical values use  / / /  keys. 
Enter "100" and press   key.
6 Press   key for two seconds to enter the RUN 
mode.