102
Function < Automatic correction of reference mass> in mode
<
Parts counting > is enabled at the moment of estimating the sample
quantity and signalled by displaying <PCS> and <SMP> (single part/piece
mass) on the top part of the display.
There are four criteria of working “Automatic correction of reference
mass”
function:
1. equilibrium should be reached,
2. quantity of parts/pieces should be increased,
3. added quantity of parts/pieces should not be greater than double
number of parts/pieces on the pan,
4. the new sample can be different from the old sample by ± 0.3
of pcs (absolute value),
If a user recognises that the sample quantity is adequate the unit mass
(single piece mass) can be saved (see ch. 26.6 of this manual)
and disable the function by pressing
.
Caution:
While the parts counting function is active the functionality of
button is
changed. Pressing
does not result in printing and saving weighments.
26.2.2. Minimum reference mass
It concerns the < Parts counting > working mode
Users before beginning the procedure of single piece mass evaluation can
declare “minimum reference mass” i.e. minimum total weight of all pieces put
on the scale pan expressed in reading divisions.
Procedure:
Enter parameter group <
Working modes> according to ch.
24 of this manual,
Choose
Parts counting / Minimum reference mass”
and then set an appropriate value.