16.4.2. Prefix / Suffix
Users can edit a prefix < Prefix> or / and suffix < Suffix> in order
to adjust the program to accept transmission frames from the scanner.
Caution:
A special protocol is required in order the code be received by RADWAG
equipment. It is required to program an appropriate prefix and suffix.
Prefix – one byte 01 hexadecimal form, suffix one byte 0D hexadecimal form.
The detailed description of cooperation scale barcode scanner can
be found in APPENDIX 06 in this manual.
Procedure:
Enter < Barcode Scanner> according to ch.16.3 of the manual,
Chose parameter <
Prefix> and then enter, using the screen
keyboard, a required value (hexadecimal) and confirm it by pressing
.
Chose parameter <
Suffix> and then enter, using the screen
keyboard, a required value (hexadecimal) and confirm it by pressing
.
16.4.3. Field selection
This option is connected with selecting data which the program will search after
reading a barcode.
Procedure:
Enter < Devices> according to ch.16 of the manual,
Chose
Barcode Scanner / Field selection and then
the following list will be displayed: