If parameter < Mode> is set to <Number> use the screen
keyboard to enter the required counter value as a threshold to trigger
off printing CC labels.
Confirm the changes introduced by pressing .
24.7. Statistics
All statistics are continuously updated after each measurement is saved
in the scale memory. Statistics can be calculated globally (does not depend
on the selected product) or separately for every product from the product
database. It can be set in parameters <
Statistics>.
Procedure:
Enter < Working modes> parameter group as described
in chapter 24 of the manual,
Enter the required working mode and choose <
Statistics>
and then set the required option.
Options:
Global
-
global statistics,
Product
-
statistics for every product.
Caution:
In case of operation with <
Statistics> set to <Product> bear in mind that
after restarting only statistics of the last weighed product are recovered.
24.8. Differential weighing
Function < Differential weighing> enables analyzing mass changes of one
or more samples. To do it start mass of a sample must be estimated, next the
sample undergoes various processes during which some components of the
sample are subtracted or added. Finally the sample is weighed again
(differential weighing). The scales calculates the difference between two
mass values (weighment I and weighment II).