92
Procedure:
Enter parameters group <
Working Modes> according to ch. 24
of this manual,
Enter the required working mode and choose:
Labelling mode /
No. of CC labels, then the editing field <No. of CC labels>
with the screen keyboard is opened,
Enter the required number of cumulative labels and confirm
by pressing
.
24.6.4. Automatic triggering of cumulative labels
Users have access to the function of automatic triggering of printing cumulative
labels after setting parameters <
Mode> and
<
Threshold>.
Procedure:
Enter < Working modes> parameter group as described
in chapter 24 of the manual,
Enter the required working mode and choose:
Labelling mode /
C label automatic triggering / Mode and then set the required
option:
None
-
Cumulative label printout is initiated by pressing
or
*
,
Mass
-
Cumulative label printout is initiated by exceeding the
value set in parameter < Threshold>. The value
is treated as the total from single weighment,
Number
-
Cumulative label printout is initiated by exceeding the
value set in parameter < Threshold>. The value
is treated as the number of single weighment.
*) Manual printing of cumulative labels can be done in two ways
depending on the button used: