37.4.2. Tarring
Syntax: T CR LF
Possible answers:
T_A CR LF
T_D CR LF
- command accepted and in progress
-
command completed
T_A CR LF
T_v CR LF
- command accepted and in progress
-
command comprehended but tare range overflow appeared
T_A CR LF
T_E CR LF
- command accepted and in progress
-
time limit for stable result exceeded
T_I CR LF
- command comprehended but cannot be executed
37.4.3. Get tare value
Syntax: OT CR LF
Reply: OT_TARA CR LF command executed
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
O T space tare space unit space CR LF
Tare - 9 characters justified to the right
Unit - 3 characters justified to the left
Caution:
Tare values are always send in adjustment unit.
37.4.4. Set tare value
Syntax: UT_TARE CR LF, where TAREtare value
Possible replies:
UT_OK CR LF
- command completed
UT_I CR LF
- command correct, but not accessible at the moment
ES CR LF
- command incorrect (e.g. incorrect tare format)