40.6. PROFIBUS board
The PUE HY10 indicator can be equipped with PROFIBUS interface (option).
This requires installation of AB6000 Anybus-IC module which is mounted in
U1 socket on universal communication modules board (385Rxxxx board).
Fig.25 Position of AB6000 module on the 385Rxxxx board
The indicator is equipped with input and output connectors. Voltage on an
output connector is 5VDC, this provides correct operation of a terminator.
The indicator features M12 5 pin B coded connectors (for PROFIBUS DP).
Connector’s pins layout:
PROFIBUS IN
(male)
Pin1 – NC
Pin2 – A
Pin3 – NC
Pin4 – B
Pin5 – NC
PROFIBUS OUT
(female)
Pin1 - +5V
Pin2 – A
Pin3 – GND
Pin4 – B
Pin5 – NC
Caution:
If the 385Rxxxx board comprises the PROFIBUS module, then simultaneous
installation of the RS 485 module on the same connector is impossible.
The modules are installed interchangeably on the same connector of the
385Rxxxx board.