37.4.23. Activate sound signal
Syntax: BP_TIME CR LF
Where: _ - space, TIME – parameter, decimal specifying how long shall the
sound last, parameter given in [ms]. Recommended range <50 - 5000>.
Possible answers:
- command carried out, BEEP sound signal activated
- command understood but not accessible at this moment
- no parameter or incorrect format
Example:
BP_350 CR LF - activate BEEP for 350 [ms]
BP_OKCR LF - BEEP activated
Notice:
1. If value greater than the permissible high limit is given, than BEEP
sound is operated for the maximum permissible amount of time.
2. BEEP sound activated via BP command is inhibited if in-course of its
activation the sound is activated by means of other device: keypad,
touch panel, proximity sensors.
37.4.24. Give available working modes
Syntax: OMI CR LF
Possible answers:
n_Mode name CR LF
OK CR LF
command carried out, accessible working modes returned
- command understood but not accessible at this moment
Mode name – parameter, working mode name, inserted in between inverted
comas. The name takes form given on a particular balance display, it is
provided in a currently selected language.
n – parameter, decimal value determining working mode number