Means of operation:
The bargraph consists of 8 red fields and three green fields.
The green fields signal weighings between MIN and MAX
threshold, where:
MIN = the minimum threshold of acceptable weighing - LO
MAX = the maximum threshold of acceptable weighing - HI
If a measurement is over the MIN (to the value of 1/3 of MIN-MAX) the
green field with a triangle on the left is visible. If the measurement is
between 1/3 and 2/3 of MIN-MAX the rectangular green field is visible.
If the measurement is between 2/3 of MIN-MAX and MAX a green field
with a triangle on the right is visible.
If the mass value is below the MIN threshold red fields with red arrows
on the left are visible. The lower mass value the more red arrows are
visible.
If the mass value is over the MAX threshold red fields with red arrows
on the right are visible. The higher mass value the more red arrows
are visible.
Thresholds MIN and MAX are on the borders between red and green
fields.
17.5.3. Bargraph “Signalling of checkweighing ranges
Settings of “Signalling checkweighing ranges” bargraph are accessible
in submenu
Display / Bargraph / “Signalling
checkweighing ranges”:
MIN, MAX thresholds working
mode
Stable signalling of MIN, MAX thresholds is
previewed on exceeding LO threshold and
reaching stable measurement result;
Unstablesignalling of MIN, MAX thresholds
is previewed on exceeding LO threshold