List of data definable for a batch number:
Name
Change of the batch number name
Template
Batch number template, allows using variables that
are intended for a printout
35.1.4. Query for number of weighings to be deleted
The user with Administrator permissions level can delete (undo) recently
performed weighings using programmable button: < Undo weighing>.
<
Reports configuration> submenu additionally allows to declare how
many weighings are to be deleted. In order to specify number of weighings that
are to be subjected to deletion process it is necessary to activate option:
<
Ask for number of weighings to be deleted>.
Procedure:
• Enter submenu: „ Reports / Reports configuration”
according to ch. 35 of this manual,
• Activate option <
Ask for number of weighings to be deleted>,
• Upon returning to home screen and pressing programmable button
< Undo weighing>, message: <Number of weighings to be
deleted> with a numeric keyboard is displayed,
• Enter appropriate value (from 1 to 10) and press button
for confirmation, message: <Number of deleted weighings: x>
is displayed, where x – number of deleted weighings.
Caution:
When:
• Some of the weighings that are to be deleted refer to any kind of report
(report on dosing, report on formula, report on PGC control, report on
vehicle scale, report on density, report on differential weighing) then the
following message is displayed <Number of deleted weighings: x.
Remaining weighings are report-related>.