37.4.10. Switch on continuous transmission in basic unit
Syntax: C1 CR LF
Possible answers:
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space unit CR LF
37.4.11. Switch off continuous transmission in basic unit
Syntax: C0 CR LF
Possible answers:
- command comprehended but cannot be executed
- command comprehended and executed
37.4.12. Switch on continuous transmission in current unit
Syntax: CU1 CR LF
Possible answers:
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF