30
14.2. Filter
This filter is intended to suppress continuous mechanical vibrations at the
cost of stabilization time.
Procedure:
Enter < Weighing> according to ch. 14 of the manual, select
<
Filter> and then set an appropriate value.
Accessible settings:
None, V.Fast, Fast, Average, Slow.
Caution:
The higher filtering level the longer stabilization time.
14.3. Autozero
The autozero function has been implemented in order to assure precise
indications. This function controls and corrects 0” indication.
While the function is active it compares the results continuously with
constant frequency. If two sequential results differ less than the declared value
of autozero range, so the scale will be automatically zeroed and the pictograms
and will be displayed.
If AUTOZERO is disabled zero is not corrected automatically. However,
in particular cases, this function can disrupt the measurement process e.g.
slow pouring of liquid or powder on the weighing pan. In this case, it is
advisable to disable the autozero function.
Procedure:
Enter < Weighing> according to ch. 14 of the manual, select
<
Autozero> and then set an appropriate value.
Accessible settings:
-
Autozero off
-
Autozero on