104
Enter a value and confirm it by pressing , then the following
message is displayed: <Put pieces: xx> (where xxthe value
entered before),
Put the declared quantity of pieces on the pan and when the result
is stable (symbol
) confirm it by pressing .
The program automatically calculates reference unit mass and
causes starting <
Parts counting >.
Caution:
The total mass of all pieces put on the weight pan cannot be greater
than the weighing range;
The total mass of all pieces put on the weight pan cannot be less than the
value declared in parameter Minimum reference mass (see section
26.2.2). If the condition above is not fulfilled the scale displays
the message <Too low sample mass>;
The mass of one unit may not be less than 0.1 of reading division
of the scale. If this condition is not fulfilled the scale displays the message:
<Too low piece mass>.
26.5. Setting the reference mass by entering single piece mass directly to
the database
After selecting a product from the product database a mass of single
piece from the field <Mass> is used.
Procedure:
Go to < Parts counting > mode and select particular product from
<Products> list.
Caution:
The selected product has to have declared unit mass (single piece mass).
26.6. Entering the unit mass to the database
The unit mass can be described a unit mass the following way:
a) Estimate the unit mass (see 26.2 and 26.3),