EasyManua.ls Logo

RADWAG PUE HY10

RADWAG PUE HY10
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
247
40. OPTIONAL EXTENSIONS MODULES
40.1. Additional I/O module
The module has been designed to improve functionality of an indicator
for additional 12 inputs and 12 outputs. The module features optoisolated
inputs and solid-state relay outputs. It enables free inputs and outputs setup
(using indicator menu). Module’s housing cover features two cable glands
through which cables are fed, which cables are 3-metre long and terminated
with stripped wires.
Caution:
Installed 12IN/12OUT module disables installation of ANxxxx analog output
module.
40.1.1. 12IN/12OUT module: technical specifications
Output parameters
Outputs number 12
Outputs types Solid-state relay
Wire cross-section 0,14 - 0,5mm
2
Maximum switching current 0,5A DC
Maximum forward voltage 30VDC, AC
Input parameters
Input number 12
Input types Optoisolated
Wire cross-section 0,14 0,5mm
2
Control voltage 5 -24V DC

Table of Contents

Other manuals for RADWAG PUE HY10