37.4.17. Read upper threshold
Syntax: OUH CR LF
Possible answers: UH_MASA CR LF - command executed
Frame format:
1 2 3 4-12 13 14 15 16 17 18 19
U H space mass space unit space CR LF
Mass - 9 characters justified to the right
Unit - 3 characters justified to the left
37.4.18. Simulation of pressing ENTER/PRINT key
Syntax: SS CR LF
Sending a command S S CR LF to a scale causes automatic saving of a
weighing record in the database and simultaneous activating a declared
printout template.
Caution:
While sending the command to a scale, all criteria for successful measurement
execution have to be met, i.e. result control, stable mass indication, etc.
37.4.19. Change platform
Syntax: PN CR LF, where Nnumber of platforms (1 to 4)
Possible answers:
PN_OK CR LF
- command completed
PN_I CR LF
- command correct, but not accessible at the moment
ES CR LF
- command incorrect (e.g. incorrect number of platforms)