Accept the message box by pressing key and return to weighing
mode.
< Setting of start mass> parameter allows to adjust
start mass of platform 1.
Caution:
Adjustment process for remaining platforms is likewise.
22.2. Start mass adjustment
It is possible to adjust only a start mass, it helps to correct the start zero
when the span does not change.
Procedure:
Enter submenu < User Adjustment > according to ch. 22
and select:
Setting of start mass”,
After entering the parameter the following message box appears:
Take the load off the pan of platform 1,
Press button . The following message appears during adjusting
start mass: „Evaluation of starting mass”,

Table of Contents