262
Status bits:
0
- correct measurement (the scale does not report an error)
1
- stable measurement
2
- scale in precise zero
3
- scale tarred
4
- scale in 2nd measuring range
5
- scale in 3rd measuring range
6
- scale reports NULL error
7
- scale reports LH error
8
- scale reports FULL error
Example:
Bit no.
B8 B7 B6 B5 B4 B3 B2 B1 B0
Value 0 0 0 0 1 0 0 1 1
The scale does not report an error, the measurement is stabilized in
the 2nd measuring range.
LOresponse is the value of LO limit in an adjustment unit of a given
weighing platform.
Process status determines status of a process
Decimal value of a
variable
Process status
Bit no.
B1
B0
0
Process inactive 0 0
1
Process start 0 1
2
Process stop 1 0
3
Process end 1 1
Inputs statusresponse is the status of set inputs
Input no.
12
11
10
9
8
7
6
5
4
3
2
1
OFF 0 0 0 0 0 0 0 0 0 0 0 0
ON 1 1 1 1 1 1 1 1 1 1 1 1