392User Manual 1179.4363.02 ─ 01
Parameters:
<Hysteresis> <numeric value>
*RST: 100 kHz
Default unit: Hz
Example: //Define a trigger hysteresis
TRIG:SOUR FREQ
TRIG:FREQ:HYST 200KHZ
Manual operation: See "Defining a trigger hysteresis" on page 135
TRIGger[:SEQuence]:HOLDoff[:TIME] <Offset>
Defines the time offset between the trigger event and the start of the measurement.
Parameters:
<Offset>
*RST: 0 s
Default unit: S
Example:
TRIG:HOLD 500us
Manual operation: See " Trigger Offset " on page 135
TRIGger[:SEQuence]:IFPower:HOLDoff <Period>
This command defines the holding time before the next trigger event.
Note that this command can be used for any trigger source, not just IF Power
(despite the legacy keyword).
Parameters:
<Period> Range: 0 s to 10 s
*RST: 0 s
Default unit: S
Example:
TRIG:SOUR EXT
Sets an external trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Manual operation: See " Trigger Holdoff " on page 135
TRIGger[:SEQuence]:LEVel[:EXTernal<1|2>] <Level>
This command defines the external trigger level.
Prerequisites for this command
Select the external trigger (TRIGger[:SEQuence]:SOURce).
Suffix:
<1|2>
.
1|2
Selects the trigger port.