Remote Control
R&S
®
FSPN
416User Manual 1179.4363.02 ─ 01
Suffix:
<n>
.
1..n
Window
<t> 1..n
Trace
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on persistence for trace 2
DISP:TRAC2:PERS ON
Manual operation: See "Displaying persistence" on page 157
DISPlay[:WINDow<n>]:TRACe<t>:RESult[:TYPE] <Resulttype>
This command selects the type of noise represented by the trace.
Suffix:
<n>
.
1..n
Window
<t> 1..n
Trace
Parameters:
<Resulttype> PN
Phase noise results.
AM
AM noise results.
PNAM
Sum of phase noise and AM noise.
Example: //Show AM noise characteristics on trace 2
DISP:TRAC2:RES AM
Manual operation: See "Selecting the displayed result" on page 156
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe] <State>
This command turns a trace on and off.
The measurement continues in the background.
Suffix:
<n>
.
Window
<w> subwindow
Not supported by all applications
<t> Trace
Example:
DISP:TRAC3 ON
Common Analysis and Display Functions