Remote Control
R&S
®
FSPN
439User Manual 1179.4363.02 ─ 01
<k> 1..n
Limit line
Return values:
<MkrIndex> 1
Limit line is compatible to the current measurement settings.
0
Limit line is not compatible to the current measurement settings.
Example: //Query limit line compatibility
CALC:LIM:NAME 'Limit'
CALC:LIM:COMP?
Usage: Query only
CALCulate<n>:LIMit<li>:COPY <Line>
This command copies a limit line.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Line> 1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Manual operation: See " Copy Line " on page 179
CALCulate<n>:LIMit<li>:DELete
This command deletes a limit line.
Suffix:
<n>
.
Window
<li> Limit line
Manual operation: See " Delete Line " on page 179
CALCulate<n>:LIMit<li>:LOWer:OFFSet <Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Common Analysis and Display Functions