Remote Control
R&S
®
FSPN
440User Manual 1179.4363.02 ─ 01
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
CALCulate<n>:LIMit<li>:LOWer:STATe <State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<li>:NAME on page 446.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See " Visibility " on page 178
CALCulate<n>:LIMit<li>:STATe <State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
on page 441).
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Common Analysis and Display Functions