EasyManua.ls Logo

R&S FSPN - Page 440

R&S FSPN
555 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FSPN
440User Manual 1179.4363.02 ─ 01
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
CALCulate<n>:LIMit<li>:LOWer:STATe <State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<li>:NAME on page 446.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See " Visibility " on page 178
CALCulate<n>:LIMit<li>:STATe <State>
This command turns the limit check for a specific limit line on and off.
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Note that a new command exists to activate the limit check and define the trace to be
checked in one step (see CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
on page 441).
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Common Analysis and Display Functions

Table of Contents